1
00:00:00,566 --> 00:00:04,333
In this lesson we will learn about oscilloscope probe calibration
2
00:00:06,700 --> 00:00:09,600
First, let's talk about oscilloscope probes
3
00:00:09,900 --> 00:00:12,833
Tektronix oscilloscope is used in this course
4
00:00:13,433 --> 00:00:16,866
This is the original probe of Tektronix oscilloscope
5
00:00:22,933 --> 00:00:27,166
There are probe compensation adjustment knobs, signal input terminals,
6
00:00:27,500 --> 00:00:31,500
ground clips and attenuation ratio adjustments on this probe
7
00:00:34,600 --> 00:00:41,266
When calibrating the oscilloscope probe, first we need to connect the oscilloscope probe to the position of CH1
8
00:00:43,733 --> 00:00:47,533
Then adjust the attenuation factor on the probe to 10 times
9
00:00:51,000 --> 00:00:54,866
It is at this position that the magnification is set to 10
10
00:00:56,400 --> 00:00:59,666
Then clamp the grounding clip on the grounding contact piece,
11
00:01:01,933 --> 00:01:05,566
and connect the input terminal to the compensation contact piece
12
00:01:08,533 --> 00:01:11,200
We observe the waveform on the screen
13
00:01:13,500 --> 00:01:17,133
If there is insufficient compensation or overcompensation,
14
00:01:17,633 --> 00:01:23,733
we can adjust it on the compensation adjustment knob with a non-inductive screwdriver until it is normal
15
00:01:28,633 --> 00:01:33,300
During the use of the oscilloscope, the probe should be calibrated regularly
16
00:01:36,566 --> 00:01:39,433
Ok, here is the oscilloscope probe calibration