1
00:00:00,300 --> 00:00:04,566
Hello everyone, today we will learn the basic settings of the oscilloscope before use
2
00:00:08,200 --> 00:00:15,066
First, connect the oscilloscope probe to the oscilloscope, that is, to its CH1 or CH2 input
3
00:00:18,833 --> 00:00:22,966
Turn the attenuation ratio adjustment switch to the position of 10
4
00:00:25,400 --> 00:00:30,633
Of course, there are also some oscilloscopes whose attenuation ratio is fixed and cannot be adjusted
5
00:00:33,533 --> 00:00:37,600
The next step is to press the auxiliary button to adjust the language
6
00:00:45,200 --> 00:00:48,300
In the second step, press the CH1 button,
7
00:00:48,500 --> 00:00:52,900
set the coupling to DC, and turn off the bandwidth limit
8
00:00:54,833 --> 00:00:57,500
Set the volts/ grid adjustment to coarse,
9
00:00:57,766 --> 00:01:00,266
and turn off the waveform inversion function
10
00:01:04,000 --> 00:01:07,433
If you turn it on, this waveform will be reversed,
11
00:01:07,500 --> 00:01:09,700
so this function is generally not used
12
00:01:10,400 --> 00:01:14,300
Next, in the third step, press the display setting button,
13
00:01:15,266 --> 00:01:17,200
set the display type to vector,
14
00:01:17,500 --> 00:01:19,766
and turn off the waveform hold function
15
00:01:21,533 --> 00:01:24,000
The format is set to YT format,
16
00:01:24,366 --> 00:01:27,366
and the contrast is generally set to about 40%
17
00:01:27,800 --> 00:01:29,966
display format is set to normal
18
00:01:32,466 --> 00:01:39,033
Step 4, press the trigger setting button to set the trigger type to edge trigger
19
00:01:44,300 --> 00:01:46,600
Set this trigger channel to CH1
20
00:01:51,200 --> 00:01:55,966
We can set the slope as rising edge or falling edge according to the measured waveform
21
00:01:58,000 --> 00:02:02,133
Usually when measuring low-level active signals, we use falling edges,
22
00:02:04,633 --> 00:02:08,933
such as signals with "#" and signals with pull-up resistors.
23
00:02:09,200 --> 00:02:12,266
When capturing power-down, use the falling edge
24
00:02:12,533 --> 00:02:14,733
Use the rising edge to measure the power supply
25
00:02:16,200 --> 00:02:19,233
There are two trigger modes: automatic and normal,
26
00:02:20,566 --> 00:02:25,333
and it can also be set as a single trigger mode by pressing the single trigger button
27
00:02:27,966 --> 00:02:30,300
The bottom coupling is set to DC
28
00:02:30,766 --> 00:02:34,866
The fifth step, according to the voltage value of the measured waveform,
29
00:02:35,466 --> 00:02:39,433
set the volts/grid, time base, source, slope,
30
00:02:39,666 --> 00:02:44,266
trigger level, horizontal position, vertical position, etc.
31
00:02:45,500 --> 00:02:50,200
For example, for a 5V inductor, we can set it to 2V per grid,
32
00:02:52,133 --> 00:02:54,966
just two and a half grids can display the waveform
33
00:02:57,966 --> 00:03:02,066
The time base can generally be set to about 500 microseconds,
34
00:03:02,900 --> 00:03:05,800
and the current setting is 250 microseconds
35
00:03:07,200 --> 00:03:09,533
Trigger channel, we choose channel 1
36
00:03:10,800 --> 00:03:13,766
The slope, which we choose to be the rising edge
37
00:03:14,500 --> 00:03:19,600
The trigger level is set to 1.8V, which is 1/3 of the measured voltage
38
00:03:20,866 --> 00:03:23,600
The trigger position is set in the middle of the screen
39
00:03:24,800 --> 00:03:28,366
The vertical position is also the position in the middle of the screen
40
00:03:28,733 --> 00:03:30,200
we call it a cross
41
00:03:31,800 --> 00:03:35,366
After all these are set, we can start to measure the waveform
42
00:03:37,633 --> 00:03:41,733
The front buttons do not need to be set every time the waveform is measured
43
00:03:46,200 --> 00:03:50,166
After setting this time, we don't need to set many functions next time,
44
00:03:50,633 --> 00:03:53,933
such as language, display type, etc.
45
00:03:54,866 --> 00:03:57,266
The next time you measure other waveforms,
46
00:03:57,533 --> 00:04:02,500
you only need to adjust the voltage of a few volts per grid, adjust the time base,
47
00:04:02,866 --> 00:04:07,200
adjust the slope and trigger level, and the others do not need to be adjusted
48
00:04:08,466 --> 00:04:11,600
Ok, this is the basic setup before using the oscilloscope