• Learning center

191. Basic setting of oscilloscope before use
We will learn the basic settings of the oscilloscope before use.
Detail
Comments

1

00:00:00,300 --> 00:00:04,566

Hello everyone, today we will learn the basic settings of the oscilloscope before use

2

00:00:08,200 --> 00:00:15,066

First, connect the oscilloscope probe to the oscilloscope, that is, to its CH1 or CH2 input

3

00:00:18,833 --> 00:00:22,966

Turn the attenuation ratio adjustment switch to the position of 10

4

00:00:25,400 --> 00:00:30,633

Of course, there are also some oscilloscopes whose attenuation ratio is fixed and cannot be adjusted

5

00:00:33,533 --> 00:00:37,600

The next step is to press the auxiliary button to adjust the language

6

00:00:45,200 --> 00:00:48,300

In the second step, press the CH1 button,

7

00:00:48,500 --> 00:00:52,900

set the coupling to DC, and turn off the bandwidth limit

8

00:00:54,833 --> 00:00:57,500

Set the volts/ grid adjustment to coarse,

9

00:00:57,766 --> 00:01:00,266

and turn off the waveform inversion function

10

00:01:04,000 --> 00:01:07,433

If you turn it on, this waveform will be reversed,

11

00:01:07,500 --> 00:01:09,700

so this function is generally not used

12

00:01:10,400 --> 00:01:14,300

Next, in the third step, press the display setting button,

13

00:01:15,266 --> 00:01:17,200

set the display type to vector,

14

00:01:17,500 --> 00:01:19,766

and turn off the waveform hold function

15

00:01:21,533 --> 00:01:24,000

The format is set to YT format,

16

00:01:24,366 --> 00:01:27,366

and the contrast is generally set to about 40%

17

00:01:27,800 --> 00:01:29,966

display format is set to normal

18

00:01:32,466 --> 00:01:39,033

Step 4, press the trigger setting button to set the trigger type to edge trigger

19

00:01:44,300 --> 00:01:46,600

Set this trigger channel to CH1

20

00:01:51,200 --> 00:01:55,966

We can set the slope as rising edge or falling edge according to the measured waveform

21

00:01:58,000 --> 00:02:02,133

Usually when measuring low-level active signals, we use falling edges,

22

00:02:04,633 --> 00:02:08,933

such as signals with "#" and signals with pull-up resistors.

23

00:02:09,200 --> 00:02:12,266

When capturing power-down, use the falling edge

24

00:02:12,533 --> 00:02:14,733

Use the rising edge to measure the power supply

25

00:02:16,200 --> 00:02:19,233

There are two trigger modes: automatic and normal,

26

00:02:20,566 --> 00:02:25,333

and it can also be set as a single trigger mode by pressing the single trigger button

27

00:02:27,966 --> 00:02:30,300

The bottom coupling is set to DC

28

00:02:30,766 --> 00:02:34,866

The fifth step, according to the voltage value of the measured waveform,

29

00:02:35,466 --> 00:02:39,433

set the volts/grid, time base, source, slope,

30

00:02:39,666 --> 00:02:44,266

trigger level, horizontal position, vertical position, etc.

31

00:02:45,500 --> 00:02:50,200

For example, for a 5V inductor, we can set it to 2V per grid,

32

00:02:52,133 --> 00:02:54,966

just two and a half grids can display the waveform

33

00:02:57,966 --> 00:03:02,066

The time base can generally be set to about 500 microseconds,

34

00:03:02,900 --> 00:03:05,800

and the current setting is 250 microseconds

35

00:03:07,200 --> 00:03:09,533

Trigger channel, we choose channel 1

36

00:03:10,800 --> 00:03:13,766

The slope, which we choose to be the rising edge

37

00:03:14,500 --> 00:03:19,600

The trigger level is set to 1.8V, which is 1/3 of the measured voltage

38

00:03:20,866 --> 00:03:23,600

The trigger position is set in the middle of the screen

39

00:03:24,800 --> 00:03:28,366

The vertical position is also the position in the middle of the screen

40

00:03:28,733 --> 00:03:30,200

we call it a cross

41

00:03:31,800 --> 00:03:35,366

After all these are set, we can start to measure the waveform

42

00:03:37,633 --> 00:03:41,733

The front buttons do not need to be set every time the waveform is measured

43

00:03:46,200 --> 00:03:50,166

After setting this time, we don't need to set many functions next time,

44

00:03:50,633 --> 00:03:53,933

such as language, display type, etc.

45

00:03:54,866 --> 00:03:57,266

The next time you measure other waveforms,

46

00:03:57,533 --> 00:04:02,500

you only need to adjust the voltage of a few volts per grid, adjust the time base,

47

00:04:02,866 --> 00:04:07,200

adjust the slope and trigger level, and the others do not need to be adjusted

48

00:04:08,466 --> 00:04:11,600

Ok, this is the basic setup before using the oscilloscope